Multimodality imaging and image guidance techniques for endovascular ascending aortic repair

Open AccessPublished:August 07, 2022DOI:https://doi.org/10.1016/j.xjtc.2022.07.024

      Key Words

      Abbreviations and Acronyms:

      2D (2-dimensional), 3D (3-dimensional), AAo (ascending aorta), ATAD (Acute type A aortic dissection), aTEVAR (ascending aorta thoracic endovascular aortic repair), CTA (computed tomography angiography), ECG (electrocardiogram), ICE (intracardiac echocardiography), IVUS (intravascular ultrasound), MPR (multiplanar view), MR (magnetic resonance), STJ (sinotubular junction), TEE (transesophageal echocardiography)
      Figure thumbnail fx1
      Different imaging modalities provide accurate depiction of the entry tear and aortic valve.
      CT, fluoroscopy, ultrasound-based imaging, and fusion techniques are complementary for intraoperative guidance in endovascular repair of the ascending aorta and can help achieve techical success.
      Endovascular ascending aortic repair has evolved from a few anecdotal cases to a viable option in patients who are not fit for surgery. Deploying stent-grafts can be challenging in this dynamic aortic segment with severe angulations, overcoming hemodynamic forces, and without impinging aortic valve leaflets. Understanding the interplay between imaging modalities is critical for technical success.
      Management of ascending aortic (AAo) pathologies such as aneurysms, pseudoaneurysms, and acute dissections have traditionally been a realm of open cardiac surgery under cardiopulmonary bypass. Recently, there has been a growing interest in endovascular treatment options for AAo pathologies.
      • Roselli E.E.
      • Atkins M.D.
      • Brinkman W.
      • Coselli J.
      • Desai N.
      • Estrera A.
      • et al.
      ARISE: first-in-human evaluation of a novel stent graft to treat ascending aortic dissection.
      • Tsilimparis N.
      • Drewitz S.
      • Detter C.
      • Spanos K.
      • von Kodolitsch Y.
      • Rohlffs F.
      • et al.
      Endovascular repair of ascending aortic pathologies with tubular endografts: a single-center experience.
      • Roselli E.E.
      • Idrees J.J.
      • Johnston D.R.
      • Eagleton M.J.
      • Desai M.Y.
      • Svensson L.G.
      Zone zero thoracic endovascular aortic repair: a proposed modification to the classification of landing zones.
      • Baikoussis N.G.
      • Antonopoulos C.N.
      • Papakonstantinou N.A.
      • Argiriou M.
      • Geroulakos G.
      Endovascular stent grafting for ascending aorta diseases.
      Ascending aortic thoracic endovascular repair (aTEVAR) has evolved from few anecdotal cases to a treatment option with excellent technical success and encouraging outcomes in patients with prohibitive risk for conventional open repair.
      • Muetterties C.E.
      • Menon R.
      • Wheatley G.H.
      A systematic review of primary endovascular repair of the ascending aorta.
      ,
      • Khoynezhad A.
      • Donayre C.E.
      • Walot I.
      • Koopmann M.C.
      • Kopchok G.E.
      • White R.A.
      Feasibility of endovascular repair of ascending aortic pathologies as part of an FDA-approved physician-sponsored investigational device exemption.
      Thoracic stent-grafts, abdominal cuffs, or custom-made grafts such as the Valiant PS-IDE (Medtronic) and Relay Non-Bare Stent (Terumo Aortic) or are most commonly used for aTEVAR in an off-label manner.
      • Muetterties C.E.
      • Menon R.
      • Wheatley G.H.
      A systematic review of primary endovascular repair of the ascending aorta.
      ,
      • Piffaretti G.
      • Grassi V.
      • Lomazzi C.
      • Brinkman W.T.
      • Navarro T.P.
      • Jenkins M.P.
      • et al.
      Thoracic endovascular stent graft repair for ascending aortic diseases.
      • Piffaretti G.
      • Czerny M.
      • Riambau V.
      • Gottardi R.
      • Wolfgruber T.
      • Probst C.
      • et al.
      Endovascular repair of ascending aortic diseases with custom-made endografts.
      • Preventza O.
      • Le Huu A.
      • Olive J.
      • Cekmecelioglu D.
      • Coselli J.S.
      Endovascular repair of the ascending aorta: the last frontier.
      Even though there are currently no devices approved by the Food and Drug Administration available, newer endovascular devices are being designed specifically for the AAo.
      • Roselli E.E.
      • Atkins M.D.
      • Brinkman W.
      • Coselli J.
      • Desai N.
      • Estrera A.
      • et al.
      ARISE: first-in-human evaluation of a novel stent graft to treat ascending aortic dissection.
      ,
      • Preventza O.
      • Le Huu A.
      • Olive J.
      • Cekmecelioglu D.
      • Coselli J.S.
      Endovascular repair of the ascending aorta: the last frontier.
      Endovascular treatment of thoracic aortic disease - full text view - ClinicalTrials.gov [Internet].
      Feasibility of endovascular repair of ascending aortic pathologies - full text view - ClinicalTrials.gov [Internet].
      The dynamic nature of AAo, with severe angulations, larger diameter with shorter lengths, proximity to coronary arteries, aortic valve leaflets, and pronounced displacement forces during cardiac cycle pose strict morphologic and anatomic criteria for device selection.
      • Belvroy V.M.
      • Zubair M.M.
      • Van Herwaarden J.A.
      • Trimarchi S.
      • Moll F.L.
      • Bismuth J.
      Important longitudinal and circumferential pulsatile changes in zone 0 of the aorta during the cardiac cycle.
      Understanding the relationship of thoracic aortic aneurysm and dissection intimal flap to the sinotubular junction (STJ) is also critical for aTEVAR. Given these factors, aTEVAR should not be approached as a conventional TEVAR procedure in AAo but with emphasis on better pre- and intraprocedural imaging. This article provides a brief overview of multimodality imaging, intraoperative real-time imaging, and image guidance techniques in the context of aTEVAR.

      Preoperative Imaging

      Step 1: Evaluation of Anatomic Suitability

      Appropriate patient selection is crucial for aTEVAR, which is typically considered in patients who are poor candidates for open repair. A few potential exclusion criteria include severe aortic insufficiency, the presence of mechanical aortic valve, and patent coronary bypass grafts arising from the AAo (Table 1). Preoperative imaging should help with determining appropriate, proximal, and distal landing zones (LZ), with adequate dimensions and distance of the proximal entry tear from STJ (∼1-2 cm). To ascertain this key information, an electrocardiogram (ECG)-gated computed tomography angiography (gated-CTA) imaging protocol must be implemented, as conventional nongated CTA scans cannot differentiate motion artifacts from entry tear of acute type A aortic dissection (ATAD) intimal flap
      • Solomon J.
      • Wilson J.
      • Samei E.
      Characteristic image quality of a third generation dual-source MDCT scanner: noise, resolution, and detectability.
      (Table 2).
      Table 1US-based early feasibility studies (EFS) and clinical trials
      ARISEEVOLVE aorta (AA arm)Medtronic valiant
      StatusEarly feasibility study (FDA IDE G140230)Clinical trial (NCT0058317)Clinical trial (NCT02201589)
      DeviceAscending Stent GraftAscending Aortic Endograft (Zenith Ascend TAA Endovascular Graft)Valiant with Captivia System
      ManufacturerWL Gore & AssociatesCook MedicalMedtronic
      Pathology
      • a.
        DeBakey type I/II AD
      • a.
        Focal aneurysm in AA
      • b.
        Pseudoaneurysms and/or dissections that are distal to the STJ
      • c.
        TAAA ≥5.0 cm in women and ≥5.5 cm in men
      • a.
        Type A AD
      • b.
        Retrograde type A AD,
      • c.
        Intramural hematoma, penetrating ulcer or pseudoaneurysm of the AA
      Device-specific inclusion criteria
       PLZ length≥20 mm distal to the most distal coronary artery ostium≥15 mm distal to the most distal coronary artery ostium≥10 mm
       PLZ diameter24-42 mm>20 mm and ≤38 mm28-44 mm
       DLZ lengthN/A>5 mm proximal to the IA≥10 mm
       DLZ diameter24-42 mm≤42 mm28-44 mm
      Clinical inclusion criteria
       Risk for open repairHighHigh or prohibitiveHigh
       Aortic valveAortic insufficiency <3+ native aortic valveN/ANo involvement in the disease
      FDA, Food and Drug Administration; IDE, Investigational Device Exemption; AD, aortic dissection; AA, ascending aorta, STJ, sinotubular junction; TAAA, thoracoabdominal aortic aneurysm; PLZ, proximal landing zone; DLZ, distal landing zone; IA, innominate artery; N/A, not applicable.
      Table 2Scanning protocols for triple rule-out CTA, aorta CTA, and retrospectively ECG-gated CTA
      ParameterTriple rule-out CTAAorta CTARetrospectively ECG-gated CTA
      TypeSpiralSpiralSpiral
      Anatomical coverageLung apices through diaphragmEntire aortic vasculatureAortic arch to cardiac apex
      Rotation time, s0.250.50.25
      Collimation, mm128 × 0.6192 × 0.6152 × 0.6
      PitchHR dependent0.6HR dependent
      Tube potential, kV100120120
      Effective mAsCareDose (ref. 288)CareDose (ref. 288)CareDose (ref. 288)
      Contrast typeNonionicNonionicNonionic
      Contrast volume, mL12012080-100
      Saline flush, mL30N/A30
      Injection rate, mL/s54-55-7
      HU trigger value230 in ascending aorta230 in abdominal aorta230 in ascending aorta
      Slice thickness1-3 mm1 mm1 mm
      Scan phases3 phasesNoncontrast, arterialArterial, 0%-100% reconstructed at every 5%-10% of R-R interval
      CTA, Computed tomography angiography; ECG, electrocardiogram; HR, Heart rate; HU, Hounsfield units.

      Step 2: Identification of Proximal LZ

      To better ascertain the proximal and distal landing zone, orthogonal multiplanar views (MPRs) of AAo are reconstructed using 3-dimensional (3D) postprocessing tools and a semi-automated aortic centerline is computed. Accuracy of aortic centerline is verified (and manually edited if necessary) by scrolling through MPR views and then reconstructed into 2-dimensional (2D) curved-MPR views for accurate length measurements.

      Step 3: 3D Planning, Sizing, and Device Selection

      Planning for aTEVAR involves assessment of vascular access to deliver the stent graft and selection of appropriate device size. Conventional CTA might miss a proximal entry tear that is mobile, especially if ECG-gating was not performed (Figure E1). While this is performed typically on CTA, real-time imaging with transesophageal echocardiography (TEE) can also be helpful to study the status of ATAD intimal flap and to correlate the measurements from CTA. This highlights the need for covisualizing the TEE images with CTA even before aTEVAR to compare and complement the information from each imaging modality. Once morphometrics are completed, the appropriate oversizing ratio depending on the pathology should be taken into consideration before device selection. In the setting of ATAD, oversizing is minimal, with most authors remaining between 0% and 5% because the treatment goal is to achieve false lumen thrombosis and eventual remodeling of the aorta without risk of rupture.
      • Muetterties C.E.
      • Menon R.
      • Wheatley G.H.
      A systematic review of primary endovascular repair of the ascending aorta.
      In the setting of aneurysmal/pseudoaneurysmal disease, the radial forces of the stent graft must allow for sturdy fixation onto the healthy aortic wall and an oversizing ratio of 20% to 30% is usually considered.
      • Roselli E.E.
      • Idrees J.J.
      • Johnston D.R.
      • Eagleton M.J.
      • Desai M.Y.
      • Svensson L.G.
      Zone zero thoracic endovascular aortic repair: a proposed modification to the classification of landing zones.
      Finally, appropriate C-arm angles are preoperatively computed based on CTA, for setting up ideal working projection for device deployment.

      Interventional Guidance

      Step 4: Vascular Access

      The optimal vascular access strategy can be planned better off CTA after assessing vessel size, calcifications, and tortuosity. In patients who cannot undergo CTA, other imaging modalities such as ferumoxytol-contrast magnetic resonance (MR) angiography can be considered.
      • Kallianos K.
      • Henry T.S.
      • Yeghiazarians Y.
      • Zimmet J.
      • Shunk K.A.
      • Tseng E.E.
      • et al.
      Ferumoxytol MRA for transcatheter aortic valve replacement planning with renal insufficiency.
      Depending on the diameter profile of stent-graft delivery system, the vascular access can be transfemoral, transapical, or transcarotid/axillary in very rare situations.
      • Roselli E.E.
      • Idrees J.J.
      • Johnston D.R.
      • Eagleton M.J.
      • Desai M.Y.
      • Svensson L.G.
      Zone zero thoracic endovascular aortic repair: a proposed modification to the classification of landing zones.
      ,
      • Muetterties C.E.
      • Menon R.
      • Wheatley G.H.
      A systematic review of primary endovascular repair of the ascending aorta.
      Although femoral access is used in most cases due to its operative ease and surgeon familiarity with it, the long distance to cover as well as iliac artery and arch tortuosity may lead to decreased trackability, pushability, and torquability of the stent-graft delivery system. Transfemoral access is the standard approach, but complications such as iliac artery rupture, dissection, and femoral pseudoaneurysms occur in up to 12.9% of cases.
      • Fernandez J.D.
      • Craig J.M.
      • Garrett H.E.
      • Burgar S.R.
      • Bush A.J.
      Endovascular management of iliac rupture during endovascular aneurysm repair.
      Iliac conduits or endovascular techniques such as “paving and cracking” for smaller diameters or “pull-down” technique for iliac kinking have been described. Transapical access provides the shorter and straighter route to the AA, but some devices have not been designed for retrograde deployment.
      • Ghazy T.G.
      • Ouda A.S.
      • Mashhour A.M.
      • Wilbring M.
      • Matschke K.
      • Kappert U.W.
      Transapical aortic stenting: an initial case series.
      Transapical access as an alternate option provides a rather direct approach to AAo and can be considered when transfemoral access is challenging. Percutaneous transapical access has been described,
      • Dudiy Y.
      • Kliger C.
      • Jelnin V.
      • Elisabeth A.
      • Kronzon I.
      • Ruiz C.E.
      Percutaneous transapical access: current status.
      but a left minithoracotomy is commonly used. Transaxillary access allows for better accommodation of the stent graft on the inner AA curvature, can be used as a cardiopulmonary bypass bailout access, and is usually less atherosclerotic compared with iliofemoral axes.
      • Dahle T.G.
      • Kaneko T.
      • McCabe J.M.
      Outcomes following subclavian and axillary artery access for transcatheter aortic valve replacement: Society of the Thoracic Surgeons/American College of Cardiology TVT Registry report.

      Step 5: Image-Fusion and Interventional Guidance

      Multimodality imaging and image guidance during interventions has gained further significance due to the recent evolution of transcatheter treatment options for valvular heart disease and complex aortic pathologies.
      • Tenorio E.R.
      • Oderich G.S.
      • Sandri G.A.
      • Ozbek P.
      • Kärkkäinen J.M.
      • Macedo T.A.
      • et al.
      Impact of onlay fusion and cone beam computed tomography on radiation exposure and technical assessment of fenestrated-branched endovascular aortic repair.
      In addition to 2D fluoroscopy, ATAD interventions can benefit when complementary information from multiple imaging modalities can be fused or covisualized to better provide interventional guidance during stent-graft deployment. For example, while aortic root could be better imaged with real-time TEE/intracardiac echocardiography (ICE) imaging, the aortic arch and visceral aorta could be better imaged with intravascular ultrasound (IVUS)/angiography.
      • Alkhouli M.
      • Hijazi Z.M.
      • Holmes D.R.
      • Rihal C.S.
      • Wiegers S.E.
      Intracardiac echocardiography in structural heart disease interventions.
      Integrating fluoroscopy with CTA after image fusion can provide broader 3D context of catheter/wire position inside aorta and its relationship to branch vessels, before deploying stent-grafts. The clinical benefits of such image fusion techniques have been reported to impact deployment precision and reduction of radiation exposure and contrast medium.
      • Goudeketting S.R.
      • Heinen S.G.H.
      • Ünlü Ç.
      • Van Den Heuvel D.A.F.
      • De Vries J.P.P.M.
      • Van Strijen M.J.
      • et al.
      Pros and cons of 3D image fusion in endovascular aortic repair: a systematic review and meta-analysis.
      CTA–fluoroscopy image fusion can be achieved by 2 approaches: (1) 3D/3D image fusion, where 3D CTA is fused with intraoperative 3D noncontrast cone-beam computed tomography; and (2) 2D/3D image fusion, where 3D CTA is fused with two 2D fluoroscopic images (anteroposterior and lateral/oblique).
      During aTEVAR procedures, landmarks such as ostia of coronary/innominate arteries, STJ, and the aortic annular plane are electronically annotated in CTA and overlaid on fluoroscopy for optimizing C-arm angulations before angiography and for interventional guidance (Figure 1). In thoracic aortic aneurysm and dissection cases, the true/false lumens and site of proximal entry tear can be annotated. Although CTA–fluoroscopy image fusion has been available from different imaging vendors, routine clinical use of this technology is still hampered by current limitations such as an additional learning curve, inaccuracies due to aortic deformation with insertion of stiff guidewires/delivery systems, and patient/respiratory/cardiac movements during the procedure.
      Figure thumbnail gr1
      Figure 1CT planning and fusion. Three-dimensional volume-rendered reconstruction of an ECG-gated CTA (A) and translucent rendering (B) of a type A aortic dissection, with an entry tear ∼2.5 cm from the sinotubular junction. Anatomic landmarks, such as the left (green asterisk) and right (red asterisk) coronary arteries, sinotubular junction (yellow arrowhead and dotted line), entry tear (yellow circle), and innominate artery ostium (white arrowhead and dotted circle) are marked and overlayed on the fluoroscopic images after CT fluoroscopy image fusion. C, C-arm angulation and image fusion markers on fluoroscopy were used as a guidance to better align and deploy the proximal end of the stent graft. CT, Computed tomography; ECG, electrocardiography; CTA, computed tomography angiography.

      Step 6: Intraprocedural Echography

      TEE is the primary real-time imaging modality used for guidance during aTEVAR. TEE can also help with device sizing and monitoring any intraprocedural complications such as aortic regurgitation, hemopericardium (Figure 2). TEE has its disadvantages, including limited imaging for distal AAo, arch and proximal descending aorta due to poor acoustic window, need for general anesthesia, and need for skilled interventional echocardiographers.
      Figure thumbnail gr2
      Figure 2Intraoperative transesophageal echocardiographic guidance. Short-axis view showing good coaptation of the aortic valve leaflets (yellow arrow) (A) and Doppler ultrasound (B) demonstrating mild aortic insufficiency after crossing the valve with a stiff wire before device deployment. C and D, Long-axis views showing the realtionship of dissection flap to the wire (yellow arrowhead indicates dissection flap; white arrow shows the wire).
      ICE is an alternative option to TEE that does not require general anesthesia. Its main advantages include real-time imaging of the aortic valve and AAo, all performed by the physician. Longitudinal views from the cavoatrial junction display the AAo and aortic valve during endograft deployment, whereas short-axis views can be used postdeployment to rule out aortic valve regurgitation and any signs of annulus rupture. Limited imaging field-of-view, additional skillsets and catheter cost, catheter oscillation due to cardiac motion, and lack of continuous wave/M modes might limit its routine use.
      IVUS is another real-time imaging modality, especially useful in the setting of aortic dissection to understand the wire/catheter position in true/false lumen, and to confirm the findings from CTA imaging or in emergent cases with suboptimal preoperative imaging studies. IVUS also can be used for intraoperative measurement of aortic lumen size and in the setting of hemodynamic instability, as in aortic rupture, when measurements from CTA may be unavailable/inaccurate. IVUS can also provide real-time imaging and assessment of dissection flap and intimal tears. IVUS platform with a stiff 0.035-inch guidewire increases pushability and trackability. IVUS catheters often assume an eccentric position and cannot maintain a stable position to guide device deployment. A coaxial steerable system with fluoroscopic image integration might mitigate these limitations.

      Step 7: Device Deployment Under Multimodality Image Guidance

      In the current setting, information from multiple imaging modalities such as TEE, ICE, computed tomography, and MR can be overlaid on fluoroscopy to guide endograft deployment over a stiff guidewire. Rapid cardiac pacing is used for attenuating heart movement, although caval ballooning or pharmacologic pacing also can be used.
      • Gottardi R.
      • Wyss T.R.
      • Van Den Berg J.C.
      • Rylski B.
      • Berger T.
      • Schmidli J.
      • et al.
      Current trends in reduction or elimination of the aortic impulse during stent-graft deployment and balloon moulding during thoracic endovascular aortic repair.
      When the correct positioning is achieved, aided by image fusion technology, the endograft is deployed. Conventionally, ballooning after stent-deployment is performed in aneurysmal pathologies but avoided in dissections due to the risk of injury to the acutely dissected, fragile aorta, and migration of the stent-graft along the aorta.
      • Gottardi R.
      • Wyss T.R.
      • Van Den Berg J.C.
      • Rylski B.
      • Berger T.
      • Schmidli J.
      • et al.
      Current trends in reduction or elimination of the aortic impulse during stent-graft deployment and balloon moulding during thoracic endovascular aortic repair.
      No complications have been described in the most current literature
      • Roselli E.E.
      • Atkins M.D.
      • Brinkman W.
      • Coselli J.
      • Desai N.
      • Estrera A.
      • et al.
      ARISE: first-in-human evaluation of a novel stent graft to treat ascending aortic dissection.
      ,
      • Roselli E.E.
      • Idrees J.J.
      • Johnston D.R.
      • Eagleton M.J.
      • Desai M.Y.
      • Svensson L.G.
      Zone zero thoracic endovascular aortic repair: a proposed modification to the classification of landing zones.
      • Baikoussis N.G.
      • Antonopoulos C.N.
      • Papakonstantinou N.A.
      • Argiriou M.
      • Geroulakos G.
      Endovascular stent grafting for ascending aorta diseases.
      • Muetterties C.E.
      • Menon R.
      • Wheatley G.H.
      A systematic review of primary endovascular repair of the ascending aorta.
      ,
      • Piffaretti G.
      • Czerny M.
      • Riambau V.
      • Gottardi R.
      • Wolfgruber T.
      • Probst C.
      • et al.
      Endovascular repair of ascending aortic diseases with custom-made endografts.
      during ballooning in the AAo.

      Step 8: Postdeployment Imaging

      A final digital subtraction angiogram is performed to evaluate stent positioning and the presence of any type 1 endoleaks (Figure 3). TEE/ICE is performed to evaluate aortic valve function.
      Figure thumbnail gr3
      Figure 3Pre- and post-TEVAR comparison. A, 3-dimensional volume-rendered reconstraction of a CTA depicting a penetrating aortic ulcer in the mid-ascending aorta (yellow arrow) (A), with its corresponding fluoroscopic view (B). Sagittal and axial view of the pre- (C, E) and postopearative CTA (D, F) after deployment of stent graft in the ascending aorta, showing complete exclusion. TEVAR, Thoracic endovascular aortic repair; CTA, computed tomography angiography.

      Follow-Up

      Our follow-up regime includes a cross-sectional imaging study at 1, 6, and 12 months and yearly thereafter. The preferred modality is CTA; however, MR is also acceptable, as it reduces radiation exposure for the patients. Follow-up is completed with a full cardiologic and echocardiographic evaluation.

      Summary

      TEVAR in zone 0 is becoming increasingly relevant. Owing to the dynamic nature of the AAo, better imaging workup is essential for pre-, intra-, and postoperative evaluation. In the near future, novel endovascular devices that can better adapt and conform to AAo will be available. In addition to newer devices, codevelopment of better real-time imaging techniques for valvular regurgitation monitoring is essential. Better integration of multimodal imaging, including but not limited to cardiac CTA, TEE, ICE, and real-time hemodynamic assessment is necessary (Figure 4). Endovascular treatment of ATAD can evolve and merge with an already maturing field of transcatheter aortic valve replacement.
      Figure thumbnail gr4
      Figure 4Imaging techniques for type A aortic dissection. Ultrasound-based techniques for visualizing the aortic root include TEE, IVUS, and ICE (illustration). The illustration depicts the anatomical positioning and field of view of the 3 different ultrasound probes. Clockwise from top to bottom, different imaging modalities illustrate the proximal entry tear of a type A aortic dissection (yellow arrow), including CTA, angiography, IVUS, and TEE. The ICE probe focuses on the aortic valve. These modalities are complementary for intraoperative guidance in endovascular repair of type A dissection. TEE, Transesophageal echocardiography; IVUS, intravascular ultrasound; ICE, intracardiac echocardiography; CTA, computed tomography angiography.

      Conflict of Interest Statement

      P.C. is a full-time Research Collaborations Manager and Senior Key Expert at Advanced Therapies Division, Siemens Medical Solutions USA, Inc. Both M.J.R. and J.B. are principal investigators for Gore ARISE Ascending aortic stent graft trial. All other authors reported no conflicts of interest.
      The Journal policy requires editors and reviewers to disclose conflicts of interest and to decline handling or reviewing manuscripts for which they may have a conflict of interest. The editors and reviewers of this article have no conflicts of interest.
      We thank Rachael Whitehead, Medical illustrator, Houston Methodist Hospital, for her kind collaboration.

      Appendix 1

      Figure thumbnail fx2
      Figure E1ECG-gated CTA. Three-dimensional volume-rendered reconstruction of a CTA depicting a focal ascending aortic dissection (A) and axial planes of an MPR passing through the proximal (green line), mid- (orange line), and distal (blue line) ascending aorta. ECG gating improves image quality and correct measuring for device sizing, as well as highlighting aortic morphology throughout the cardiac cycle. On the left, measurements are taken in the systolic (B, C, D) and diastolic (B', C', D') phases, corresponding to 45% and 70% of the R'R interval. ECG, Electrocardiogram; CTA, computed tomography angiography; MPR, multiplanar reconstruction.

      References

        • Roselli E.E.
        • Atkins M.D.
        • Brinkman W.
        • Coselli J.
        • Desai N.
        • Estrera A.
        • et al.
        ARISE: first-in-human evaluation of a novel stent graft to treat ascending aortic dissection.
        J Endovasc Ther. May 19, 2022; ([Epub ahead of print])
        • Tsilimparis N.
        • Drewitz S.
        • Detter C.
        • Spanos K.
        • von Kodolitsch Y.
        • Rohlffs F.
        • et al.
        Endovascular repair of ascending aortic pathologies with tubular endografts: a single-center experience.
        J Endovasc Ther. 2019; 26: 439-445
        • Roselli E.E.
        • Idrees J.J.
        • Johnston D.R.
        • Eagleton M.J.
        • Desai M.Y.
        • Svensson L.G.
        Zone zero thoracic endovascular aortic repair: a proposed modification to the classification of landing zones.
        J Thorac Cardiovasc Surg. 2018; 155: 1381-1389
        • Baikoussis N.G.
        • Antonopoulos C.N.
        • Papakonstantinou N.A.
        • Argiriou M.
        • Geroulakos G.
        Endovascular stent grafting for ascending aorta diseases.
        J Vasc Surg. 2017; 66: 1587-1601
        • Muetterties C.E.
        • Menon R.
        • Wheatley G.H.
        A systematic review of primary endovascular repair of the ascending aorta.
        J Vasc Surg. 2018; 67: 332-342
        • Khoynezhad A.
        • Donayre C.E.
        • Walot I.
        • Koopmann M.C.
        • Kopchok G.E.
        • White R.A.
        Feasibility of endovascular repair of ascending aortic pathologies as part of an FDA-approved physician-sponsored investigational device exemption.
        J Vasc Surg. 2016; 63: 1483-1495
        • Piffaretti G.
        • Grassi V.
        • Lomazzi C.
        • Brinkman W.T.
        • Navarro T.P.
        • Jenkins M.P.
        • et al.
        Thoracic endovascular stent graft repair for ascending aortic diseases.
        J Vasc Surg. 2019; 70: 1384-1389.e1
        • Piffaretti G.
        • Czerny M.
        • Riambau V.
        • Gottardi R.
        • Wolfgruber T.
        • Probst C.
        • et al.
        Endovascular repair of ascending aortic diseases with custom-made endografts.
        Eur J Cardiothorac Surg. 2021; 59: 741-749
        • Preventza O.
        • Le Huu A.
        • Olive J.
        • Cekmecelioglu D.
        • Coselli J.S.
        Endovascular repair of the ascending aorta: the last frontier.
        Ann Cardiothorac Surg. 2022; 11: 26-30
      1. Endovascular treatment of thoracic aortic disease - full text view - ClinicalTrials.gov [Internet].
      2. Feasibility of endovascular repair of ascending aortic pathologies - full text view - ClinicalTrials.gov [Internet].
        • Belvroy V.M.
        • Zubair M.M.
        • Van Herwaarden J.A.
        • Trimarchi S.
        • Moll F.L.
        • Bismuth J.
        Important longitudinal and circumferential pulsatile changes in zone 0 of the aorta during the cardiac cycle.
        Eur J Cardiothorac Surg. 2020; 59: 467-472
        • Solomon J.
        • Wilson J.
        • Samei E.
        Characteristic image quality of a third generation dual-source MDCT scanner: noise, resolution, and detectability.
        Med Phys. 2015; 42: 4941-4953
        • Kallianos K.
        • Henry T.S.
        • Yeghiazarians Y.
        • Zimmet J.
        • Shunk K.A.
        • Tseng E.E.
        • et al.
        Ferumoxytol MRA for transcatheter aortic valve replacement planning with renal insufficiency.
        Int J Cardiol. 2017; 231: 255-257
        • Fernandez J.D.
        • Craig J.M.
        • Garrett H.E.
        • Burgar S.R.
        • Bush A.J.
        Endovascular management of iliac rupture during endovascular aneurysm repair.
        J Vasc Surg. 2009; 50: 1293-1300
        • Ghazy T.G.
        • Ouda A.S.
        • Mashhour A.M.
        • Wilbring M.
        • Matschke K.
        • Kappert U.W.
        Transapical aortic stenting: an initial case series.
        EuroIntervention. 2016; 12: 1305-1310
        • Dudiy Y.
        • Kliger C.
        • Jelnin V.
        • Elisabeth A.
        • Kronzon I.
        • Ruiz C.E.
        Percutaneous transapical access: current status.
        EuroIntervention. 2014; 10: U84-U89
        • Dahle T.G.
        • Kaneko T.
        • McCabe J.M.
        Outcomes following subclavian and axillary artery access for transcatheter aortic valve replacement: Society of the Thoracic Surgeons/American College of Cardiology TVT Registry report.
        JACC Cardiovasc Interv. 2019; 12: 662-669
        • Tenorio E.R.
        • Oderich G.S.
        • Sandri G.A.
        • Ozbek P.
        • Kärkkäinen J.M.
        • Macedo T.A.
        • et al.
        Impact of onlay fusion and cone beam computed tomography on radiation exposure and technical assessment of fenestrated-branched endovascular aortic repair.
        J Vasc Surg. 2019; 69: 1045-1058.e3
        • Alkhouli M.
        • Hijazi Z.M.
        • Holmes D.R.
        • Rihal C.S.
        • Wiegers S.E.
        Intracardiac echocardiography in structural heart disease interventions.
        JACC Cardiovasc Interv. 2018; 11: 2133-2147
        • Goudeketting S.R.
        • Heinen S.G.H.
        • Ünlü Ç.
        • Van Den Heuvel D.A.F.
        • De Vries J.P.P.M.
        • Van Strijen M.J.
        • et al.
        Pros and cons of 3D image fusion in endovascular aortic repair: a systematic review and meta-analysis.
        J Endovasc Ther. 2017; 24: 595-603
        • Gottardi R.
        • Wyss T.R.
        • Van Den Berg J.C.
        • Rylski B.
        • Berger T.
        • Schmidli J.
        • et al.
        Current trends in reduction or elimination of the aortic impulse during stent-graft deployment and balloon moulding during thoracic endovascular aortic repair.
        Eur J Cardiothorac Surg. 2021; 60: 1466-1474